Company: GE Healthcare, LLC
Date of Enforcement Report: 10/23/2019
Class II
PRODUCT
ApexPro Telemetry Server System. Also identified as Modification To: ApexPro Telemetry System and ApexPro FH Telemetry System – Product Usage: These systems are typically deployed in sub acute care areas in hospitals or clinical sites where patient mobility can enhance the effectiveness of the medical procedures administered.
- Recall Number: Z-0175-2020
REASON
May not provide visual and/or audible alarms at the CARESCAPE Central Station or Clinical Information Center monitor for ECG arrhythmias, ECG LEADS FAIL or Pulse Oximetry (SpO2) under certain conditions.
RECALLING FIRM/MANUFACTURER
GE Healthcare, LLC on 08/30/2019. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
21,132 (17,218 US; 3,914 OUS) devices total
DISTRIBUTION
Nationwide and International