Recall – Pressing and holding Stop button leads to restart of movement

Company: Philips Medical Systems Nederlands
Date of Enforcement Report: 1/8/2019
Class II

PRODUCT

Ingenia Elition S, Ingenia Elition X

Recall Number: Z-0694-2019

REASON

During automatic tabletop movement, pressing and holding the Stop Table button may lead to an unexpected restart of the table movement.

RECALLING FIRM/MANUFACTURER

Philips Medical Systems Nederlands, Best, Netherlands on 12/3/2018. Voluntary: Firm Initiated, recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
35

DISTRIBUTION
U.S. Nationwide and International