Company: Tandem Diabetes Care, Inc.
Date of Enforcement Report: 9/29/2026
Class: II
PRODUCT:
Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No
Recall Number:
Z-3330-2026
REASON:
A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM’s glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.
RECALLING FIRM/MANUFACTURER:
Tandem Diabetes Care, Inc.
VOLUME:
177
DISTRIBUTION:
US Nationwide distribution.