Company: Abiomed, Inc.
Date of Enforcement Report: 9/21/2026
Class: II
PRODUCT
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US. 7. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US.
Recall Number:
Z-3251-2026
REASON:
Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.
RECALLING FIRM/MANUFACTURER:
Abiomed, Inc.
VOLUME:
3,322 units
DISTRIBUTION:
Worldwide – US Nationwide and the countries of Austria, Belgium, Canada, France, Germany, Hong Kong, Italy, Luxembourg, Poland, Spain, Switzerland, United Kingdom.