FDA Recall: Pump unexpectedly stops delivering insulin

Company: Medtronic MiniMed, Inc.
Date of Enforcement Report: 7/8/2026
Class: II

PRODUCT:

MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896

Recall Number:

Z-2698-2026

REASON:

A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).

RECALLING FIRM/MANUFACTURER:

Medtronic MiniMed, Inc.

VOLUME:

778

DISTRIBUTION:

International distribution to the countries of Spain, Great Britain, Netherlands, Italy.

About the author

Allison joined SoftwareCPR in 2022, bringing along over ten years of passion for building teamwork and growth in organizations. This is established in part by the greatest University the South has to offer- Mississippi State.

Upcoming Public Courses

No public courses are planned at this time.

If you would like to discuss a private, onsite course, please fill out the form below.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.