FDA Recall: Pump unexpectedly stops delivering insulin

Company: Medtronic MiniMed, Inc.
Date of Enforcement Report: 7/8/2026
Class: II

PRODUCT:

MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896

Recall Number:

Z-2698-2026

REASON:

A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).

RECALLING FIRM/MANUFACTURER:

Medtronic MiniMed, Inc.

VOLUME:

778

DISTRIBUTION:

International distribution to the countries of Spain, Great Britain, Netherlands, Italy.