FDA Recall: Nine Identified Software Issues Leading to Various Issues

Company: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Date of Enforcement Report: 1/14/2026
Class: II

PRODUCT:

Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Number: 722229. 2. Azurion 3 M15. Model Number: 722230. 3. Azurion 5 M12. Model Number: 722231. 4. Azurion 5 M20. Model Number: 722232. 5. Azurion 7 B12. Model Number: 722235. 6. Azurion 7 B20. Model Number: 722236. 7. Azurion 7 M12. Model Number: 722233. 8. Azurion 7 M20. Model Number: 722234

Recall Number:

Z-1079-2026

REASON:

Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.

RECALLING FIRM/MANUFACTURER:

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

VOLUME:

990 units

DISTRIBUTION:

Worldwide distribution – US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Chile, Colombia, Cote de’Ivore, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Kosovo, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Palestine, Philippines, Poland, Portugal, R¿union, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.

About the author

Allison joined SoftwareCPR in 2022, bringing along over ten years of passion for building teamwork and growth in organizations. This is established in part by the greatest University the South has to offer- Mississippi State.

SoftwareCPR Training Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software

ISO 14971 Risk Management (typically with heavy emphasis on software aspect)

IEC 81001-5-1 Cybersecurity Lifecycle

ISO 13485 QMS Requirements & optionally ISO 19011 for leading audits

Being Agile Yet Compliant (AAMI TIR 45)

IEC 29119-1 Software and System Testing

US Classification for Software Devices

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.