FDA Recall: Software issue may result in over- or under-delivery

Company: ICU Medical, Inc.
Date of Enforcement Report: 12/19/2025
Class: II

PRODUCT

LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02

Recall Number:

Z-0966-2026

REASON:

Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.

RECALLING FIRM/MANUFACTURER:

ICU Medical, Inc.

VOLUME:

12 customers

DISTRIBUTION:

US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.

About the author

Allison joined SoftwareCPR in 2022, bringing along over ten years of passion for building teamwork and growth in organizations. This is established in part by the greatest University the South has to offer- Mississippi State.

SoftwareCPR Training Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software

ISO 14971 Risk Management (typically with heavy emphasis on software aspect)

IEC 81001-5-1 Cybersecurity Lifecycle

ISO 13485 QMS Requirements & optionally ISO 19011 for leading audits

Being Agile Yet Compliant (AAMI TIR 45)

IEC 29119-1 Software and System Testing

US Classification for Software Devices

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.