FDA Recall: Software bug could compromise diagnostic accuracy

Company: DICOM Grid, Inc.
Date of Enforcement Report: 12/18/2025
Class: II

PRODUCT

Intelerad InteleShare software, with ProViewer component

Recall Number:

Z-0941-2026

REASON:

Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are met: Viewing images in InteleShare viewer; Multiplanar reconstruction applied; Manual rotation applied, could result in inaccurate length measurement tool values that could compromise diagnostic accuracy, lead to misdiagnosis or inappropriate follow-up.

RECALLING FIRM/MANUFACTURER:

DICOM Grid, Inc.

VOLUME:

924

DISTRIBUTION:

US: AZ, TX, GA, NV, MO, MI, IL, OH, CT, OR, FL, NY, NM, MN, PA, AK, ND, NJ, VA, WA, NC, TN, WI, GA, AK, CA, SD, CO, MA, WV, MN, KS, MD, VT, IA, RI, AL, ID, LA, PR, NH, IN, OK, UT, NE, SC, AR, DE OUS: CA, NZ, AU, GB, BE, FR, IT, NO, SE

About the author

Allison joined SoftwareCPR in 2022, bringing along over ten years of passion for building teamwork and growth in organizations. This is established in part by the greatest University the South has to offer- Mississippi State.

SoftwareCPR Training Courses

ISO13485:2016 ISO 13485 Internal Audit(or) Training Course (Live, 3-day)

IEC 62304 and other Emerging Standards Impacting Medical Device Software (Live, 3-day)

Being Agile & Yet CompliantISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.