FDA Recall: Software Defect Causing Prescription Errors

Company: Vantive US Healthcare LLC
Date of Enforcement Report: 10/29/2025
Class: II

PRODUCT

Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (US, EMEA, and LATAM)

Recall Number:

Z-0409-2026

REASON:

Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to not be properly saved on the actual cycler that the patient is using to perform their therapy. As a result, the clinician may think that the patient is receiving the correct prescribed therapy because of what is erroneously displayed on the clinician’s remote Sharesource application, when in fact the cycler that is with the patient is actually delivering an outdated and incorrect therapy.

RECALLING FIRM/MANUFACTURER:

Vantive US Healthcare LLC

VOLUME:

1

DISTRIBUTION:

Worldwide distribution – US Nationwide and the countries of Australia, Hong Kong, Korea, New Zealand, Taiwan, Thailand, France, Germany, Ireland, Netherlands, Italy, Spain, Turkey, United Kingdom, Chile, and Colombia.

About the author

Allison joined SoftwareCPR in 2022, bringing along over ten years of passion for building teamwork and growth in organizations. This is established in part by the greatest University the South has to offer- Mississippi State.

Upcoming Public Courses

No public courses are planned at this time.

If you would like to discuss a private, onsite course, please fill out the form below.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.