FDA Recall: Software Defect Causing Prescription Errors

Company: Vantive US Healthcare LLC
Date of Enforcement Report: 10/29/2025
Class: II

PRODUCT

Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (US, EMEA, and LATAM)

Recall Number:

Z-0409-2026

REASON:

Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to not be properly saved on the actual cycler that the patient is using to perform their therapy. As a result, the clinician may think that the patient is receiving the correct prescribed therapy because of what is erroneously displayed on the clinician’s remote Sharesource application, when in fact the cycler that is with the patient is actually delivering an outdated and incorrect therapy.

RECALLING FIRM/MANUFACTURER:

Vantive US Healthcare LLC

VOLUME:

1

DISTRIBUTION:

Worldwide distribution – US Nationwide and the countries of Australia, Hong Kong, Korea, New Zealand, Taiwan, Thailand, France, Germany, Ireland, Netherlands, Italy, Spain, Turkey, United Kingdom, Chile, and Colombia.

About the author

Allison joined SoftwareCPR in 2022, bringing along over ten years of passion for building teamwork and growth in organizations. This is established in part by the greatest University the South has to offer- Mississippi State.

SoftwareCPR Training Courses

ISO13485:2016 ISO 13485 Internal Audit(or) Training Course (Live, 3-day)

IEC 62304 and other Emerging Standards Impacting Medical Device Software (Live, 3-day)

Being Agile & Yet CompliantISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.