FDA Recall: May experience timing variations between software tasks

Company: mo-Vis BVBA
Date of Enforcement Report: 9/2/2025
Class: I

PRODUCT

Micro Joystick R-net. Electrical wheelchair component.

Recall Number:

Z-2454-2025

REASON:

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

RECALLING FIRM/MANUFACTURER:

mo-Vis BVBA

VOLUME:

3 units

DISTRIBUTION:

US Nationwide distribution in the state of TX.