Company: mo-Vis BVBA
Date of Enforcement Report: 9/2/2025
Class: I
PRODUCT
Micro Joystick R-net. Electrical wheelchair component.
Recall Number:
Z-2454-2025
REASON:
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
RECALLING FIRM/MANUFACTURER:
mo-Vis BVBA
VOLUME:
3 units
DISTRIBUTION:
US Nationwide distribution in the state of TX.