FDA Recall: A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available

Company: Medtronic, Inc.
Date of Enforcement Report: 7/17/2025
Class: II

PRODUCT

Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D00U003; 2. Percepta Serena Solara application, Software Model Number D00U004; 3. Cobalt Crome application, Software Model Number D00U005; 4. Micra VR application, Software Model Number D00U006; 5. Micra AV application, Software Model Number D00U007; 6. Viva Consulta Syncra Advisa Ensura application, Software Model Number D00U008; 7. Claria Amplia Compia application, Software Model Number D00U009; 8. Evera MRI application, Software Model Number D00U010; 9. Visia AF application, Software Model Number D00U011; 10. Viva Brava Evera application, Software Model Number D00U012; 11. Micra VR2 AV2 application, Software Model Number D00U022

Recall Number:

Z-2127-2025

REASON:

A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device Manager. The erroneous message indicates an electrical reset has occurred. However, no reset has actually occurred and there are no parameter changes or loss of diagnostics. Prior to this software update, if an erroneous SmartSync message was displayed, confusion regarding the message could have caused a clinician to consider an unnecessary system revision. Medtronic has received 39 reports of this display error with two instances of unnecessary device explant through 30 April 2025

RECALLING FIRM/MANUFACTURER:

Medtronic, Inc.

VOLUME:

3909 units

DISTRIBUTION:

About the author

Allison joined SoftwareCPR in 2022, bringing along over ten years of passion for building teamwork and growth in organizations. This is established in part by the greatest University the South has to offer- Mississippi State.

SoftwareCPR Training Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software

ISO 14971 Risk Management (typically with heavy emphasis on software aspect)

IEC 81001-5-1 Cybersecurity Lifecycle

ISO 13485 QMS Requirements & optionally ISO 19011 for leading audits

Being Agile Yet Compliant (AAMI TIR 45)

IEC 29119-1 Software and System Testing

US Classification for Software Devices

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.