FDA Recall: Software Bug Will Have Inaccurate Calculations from PET/CT Studies Acquired on Daylight Savings Time

Company: INTELERAD MEDICAL SYSTEMS INCORPORATED
Date of Enforcement Report: 2/25/2025
Class II

PRODUCT

IntelePACS (Image Fusion Module) – InteleViewer

Recall Number:

Z-1208-2025

REASON:

A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.

RECALLING FIRM/MANUFACTURER:

INTELERAD MEDICAL SYSTEMS INCORPORATED

VOLUME OF PRODUCT IN COMMERCE:

88

DISTRIBUTION:

Worldwide – US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia

About the author

Allison joined SoftwareCPR in 2022, bringing along over ten years of passion for building teamwork and growth in organizations. This is established in part by the greatest University the South has to offer- Mississippi State.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.