FDA Recall: Abbott Issues Correction for HeartMate LVAS System Monitor due to Screen Issues that May Cause Unintentional Pump Stop

Company: Abbott Medical Group
Date of Enforcement Report: 6/7/2024
Class I

PRODUCT

Thoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN, 1286INT and L1286INT. Left ventricular assist device

Recall Number:

Z-1921-2024

REASON:

System monitor screen may display atypical behavior: Overlapping screens/buttons, Frozen screen, Distorted text, blanks, or zeroes in place of values, Unresponsive buttons where the user is unable to initiate a command, which may lead to anxiety /inconvenience, hemodynamic compromise, extended surgical time, and Pump Stop command may be selected.

RECALLING FIRM/MANUFACTURER:

Thoratec Corp.

VOLUME OF PRODUCT IN COMMERCE:

4,842

DISTRIBUTION:

Worldwide distribution. US nationwide including Puerto Rico, UNITED ARAB EMIRATES, AUSTRIA, BELGIUM, BULGARIA, SWITZERLAND, CYPRUS, CZECH REPUBLIC, GERMANY, DENMARK, ESTONIA, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, GUADELOUPE, GREECE, IRELAND, ISRAEL, IRAN, ISLAMIC REPUBLIC OF, ITALY, KUWAIT, LEBANON, LITHUANIA, LUXEMBOURG, MACEDONIA, MARTINIQUE, NETHERLANDS, NORWAY, OMAN, POLAND, PALESTINE, PORTUGAL, QATAR, ROMANIA, SERBIA, RUSSIAN FEDERATION, SWEDEN, SLOVENIA, TUNISIA, TURKEY, SOUTH AFRICA, CANADA, AUSTRALIA, HONG KONG, MALAYSIA, SINGAPORE, SOUTH KOREA, TAIWAN, THAILAND, JAPAN, BANGLADESH, INDIA, PAKISTAN, ARGENTINA, CAYMAN ISLANDS, CHILE, COLOMBIA, COSTA RICA, MEXICO, and BRAZIL.

About the author

Allison joined SoftwareCPR in 2022, bringing along over ten years of passion for building teamwork and growth in organizations. This is established in part by the greatest University the South has to offer- Mississippi State.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.