FDA Recall: Overestimated GPR Can Show False Positive Results

Company: PTW-FREIBURG
Date of Enforcement Report: 2/22/2024
Class II

PRODUCT

Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets and treatment plans in radiation oncology.

Recall Number:

Z-1181-2024

REASON:

Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation can therefore show false positive results.

RECALLING FIRM/MANUFACTURER:

PTW-FREIBURG

VOLUME OF PRODUCT IN COMMERCE:

1 US; 65 worldwide

DISTRIBUTION:

GA