FDA Recall: Software Crash when Simultaneous Use of Device Functions

Company: Philips North America
Date of Enforcement Report: 11/15/2023
Class II

PRODUCT

Spectral CT on Rails, Software Version 5.1.0, Model Number 728334

Recall Number:

Z-0306-2024

REASON:

When preparing for CCT Fluoroscopy mode and activating the CCT foot pedal simultaneously, the operator may experience a software crash. This would result in the system needing to be restarted.

RECALLING FIRM/MANUFACTURER:

Philips North America

VOLUME OF PRODUCT IN COMMERCE:

1 unit

DISTRIBUTION:

US Nationwide distribution in the state of MN.