Recall- Software Error Can Lead to Temporary Cessation of Medication Flow

Company: Fresenius Kabi USA, LLC
Date of Enforcement Report: 10/6/2022
Class I

PRODUCT

Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004 version 5.2.0

Recall Number:

Z-0002-2023

REASON

The display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does not indicate an alarm condition or respond. This issue leads to temporary cessation of flow of a medication or fluid, requiring immediate intervention by the clinician.

RECALLING FIRM/MANUFACTURER

Fresenius Kabi USA, LLC

VOLUME OF PRODUCT IN COMMERCE

4 instances

DISTRIBUTION

Domestic distribution to NJ and WI.