Recall- Software Issue May Lead to Incorrect Positions for X-Rays & Radiographs

Company: Zap Surgical Systems

Date of Enforcement Report: 9/14/2022

Class II

PRODUCT

ZAP-X Radiosurgery System

Recall Number:

Z-1742-2022

REASON

Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.

RECALLING FIRM/MANUFACTURER

Zap Surgical Systems

VOLUME OF PRODUCT IN COMMERCE

8

DISTRIBUTION

Worldwide distribution – US Nationwide distribution in the states of AZ, MD, FL, CO and the countries of DEU, CHE, JPN.