Recall- Software Issue May Lead to Incorrect Positions for X-Rays & Radiographs

Company: Zap Surgical Systems

Date of Enforcement Report: 9/14/2022

Class II

PRODUCT

ZAP-X Radiosurgery System

Recall Number:

Z-1742-2022

REASON

Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.

RECALLING FIRM/MANUFACTURER

Zap Surgical Systems

VOLUME OF PRODUCT IN COMMERCE

8

DISTRIBUTION

Worldwide distribution – US Nationwide distribution in the states of AZ, MD, FL, CO and the countries of DEU, CHE, JPN.

About the author

Allison joined SoftwareCPR in 2022, bringing along over ten years of passion for building teamwork and growth in organizations. This is established in part by the greatest University the South has to offer- Mississippi State.

Upcoming Public Courses

No public courses are planned at this time.

If you would like to discuss a private, onsite course, please fill out the form below.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.