Recall – Radiographic system processes incorrect position values

Company: Siemens Medical Solutions USA, Inc
Date of Enforcement Report: 4/6/2022
Class II

PRODUCT

Ysio system with VC10 software version, Model Number 10281013. Radiographic system

Recall Number:

Z-0841-2022

REASON

Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If the user activates a longitudinal or transversal movement at the time of the error, the system controls may initiate a correction of the tube stand position, resulting in an unintended movement. If the described error occurs, a potential for collision risk for objects or persons in the vicinity of the moving stand exists. This may result in minor to serious injuries to persons.

RECALLING FIRM/MANUFACTURER

Siemens Medical Solutions USA, Inc on 2/3/2022. Voluntary: Firm initiated

VOLUME OF PRODUCT IN COMMERCE

70 Systems

DISTRIBUTION

U.S. Nationwide

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.