Recall – Unexpected image set behavior while cine loop running

Company: RAYSEARCH LABORATORIES AB
Date of Enforcement Report: 3/16/2022
Class II

PRODUCT

RayStation/RayPlan- designed for treatment planning and analysis of radiation therapy.

Software Version: 6.0.0.24 , 6.1.0.26 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.61 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 .

Including Service Paks: Product name (build number) UDI-DI RayStation 6/RayPlan 2 (6.0.0.24) 07350002010013 RayStation 6/RayPlan 2 Service Pack 1 (6.1.1.2) 07350002010082 RayStation 6/RayPlan 2 Service Pack 2 (6.2.0.7) 07350002010075 RayStation 6/RayPlan 2 Service Pack 3 (6.3.0.6) 07350002010242 RayStation/RayPlan 7 (7.0.0.19) 07350002010068 RayStation/RayPlan 8A (8.0.0.61) 07350002010112 RayStation/RayPlan 8A Service Pack 1 (8.0.1.10) 07350002010136 RayStation/RayPlan 8B (8.1.0.47) 07350002010129 RayStation/RayPlan 8B Service Pack 1 (8.1.1.8) 07350002010204 RayStation/RayPlan 8B Service Pack 2 (8.1.2.5) 07350002010235 RayStation/RayPlan 9A (9.0.0.113) 07350002010174 RayStation/RayPlan 9B (9.1.0.933) 07350002010266 RayStation/RayPlan 9B Service Pack 1 (9.2.0.483) 07350002010297

Recall Number: Z-0732-2022

REASON

If a new primary image set is selected while the cine loop is running, the primary image set will be displayed as both primary and secondary image set in all side-by-side views. This will also be true for any new patient or case opened while the cine loop is running.

RECALLING FIRM/MANUFACTURER

RAYSEARCH LABORATORIES AB on 12/30/2021. Voluntary: Firm initiated

VOLUME OF PRODUCT IN COMMERCE

75 Licenses

DISTRIBUTION

U.S. Nationwide

Upcoming SoftwareCPR Training Courses:

Public Course – Oct 18-21, 2022 – Being Agile & Yet Compliant (virtual)

Early Bird Discount Registration through September 30, 2022.  Reserve your spot!

Register here: https://events.eventzilla.net/e/october-2022-softwarecpr-agile-and-compliant-training-course-2138573767

 

Public Course – Jan 9-11, 2023 – Risk Management (in-person)

Our newly updated ISO 14971:2019 Medical Device Risk Management, A Software Organization’s Perspective public training course is now open for registration!

Where:  Tampa, Florida

  • Coverage of ISO 14971:2019, IEC 62304; amd1, and IEC/TR 80002-1.
  • System level hazards analysis – mapping to software, cybersecurity, and usability
  • Why FMEA is incomplete for medical device risk management.
  • How to perform software hazards analysis.
  • And more!

3-days onsite with group exercises, quizzes, examples, Q&A.

Early Bird Discount Registration through September 30, 2022.  Reserve your spot!

Register here: https://events.eventzilla.net/e/2023-softwarecpr-public-training-course–iso-14971-medical-device-risk-management-a-software-organizations-perspective-2138576610

 

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.