Company: Elekta Inc. Date of Enforcement Report: 10/6/2021 Class II PRODUCT Elekta Monaco – Product Usage: used to make treatment plans for patients with prescriptions for external beam radiation therapy. Recall Number: Z-2563-2021 REASON Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to...
Read More
Company: Biodex Medical Systems, Inc. Date of Enforcement Report: 10/6/2021 Class II PRODUCT Atomlab 500 Dose Calibrator Software 2.0.00 through Revision 2.0.08, inclusive Model: 086-330. Recall Number: Z-2580-2021 REASON When deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have...
Read More