Recall – Possibility of incorrect calculated radiation treatment plan

Company: Siemens Medical Solutions USA, Inc
Date of Enforcement Report: 7/29/2020
Class II

PRODUCT

Syngo.via RT Image Suite with software versions syngo.via VB30 or VB40

Recall Number: Z-2688-2020

REASON

If the user modifies for any reason (e.g. reduction of artifacts) the original image orientation of a standard MR protocol to acquire images in a different orientation for further processing in Synthetic CT , the software does not recognize the adapted acquisition plane. This may result in images with wrong geometry. When this distortion remains unnoticed and the images are subsequently exported to a treatment planning system (TPS), an incorrect calculated radiation treatment plan cannot be excluded. The occurrence of this issue is very unlikely and has never been reported so far.

RECALLING FIRM/MANUFACTURER

Siemens Medical Solutions USA, Inc on 6/24/2020. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

21 units

DISTRIBUTION

U.S. Nationwide

Upcoming Public Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software training course in Boise, Idaho. 3-Day course.

October 13-15, 2026

Limited Early Bird Registration: $995 / person
Non refundable but transferable.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.