Recall – Patient data module negatively affected by AED

Company: GE Healthcare, LLC
Date of Enforcement Report: 7/1/2020
Class II

PRODUCT

CARESCAPE Patient Data Module (PDM) used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro

Recall Number: Z-2424-2020

REASON

The CARESCAPE Patient Data Module (PDM) is used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro. If Pace detection is turned ON, on the monitor, and if an Automated External Defibrillator (AED) is used to perform defibrillation, the low frequency signals deployed by the AED, cause the PDM to issue a false flat line ECG waveform and alarm Asystole to be displayed on the monitor.

RECALLING FIRM/MANUFACTURER

GE Healthcare, LLC on 5/18/2020. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

70,918 devices

DISTRIBUTION

International

What’s New

FDA:  Exemptions, Variances, and Alternate Forms of Adverse Event Reporting for Medical Devices

Manufacturers, importers, and user facilities can request an exemption or variance from any or all of the medical device reporting requirements in 21 CFR Part 803.  Read more:  https://www.fda.gov/medical-devices/medical-device-reporting-mdr-how-report-medical-device-problems/exemptions-variances-and-alternate-forms-adverse-event-reporting-medical-devices