Recall – Lab management system sw may display erroneous patient results

Company: Siemens Healthcare Diagnostics, Inc.
Date of Enforcement Report: 7/8/2020
Class II

PRODUCT

CentraLink Data Management System software version 16.0.2 Siemens Material Number (SMN): 11313246 (2) CentraLink Data Management System software version 16.0.3 Siemens Material Number (SMN): 11314337 Product Usage: are clinical laboratory information modules which consolidate test orders and test results, provide test ordering and priority routing information to automation, apply rules to test orders and auto-verification rules for test results, and provide integrated management of quality control of test system.

Recall Number: Z-2437-2020

REASON

Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review

RECALLING FIRM/MANUFACTURER

Siemens Healthcare Diagnostics, Inc. on 7/20/2020. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

186 units

DISTRIBUTION

International

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  TBD

Email training@softwarecpr.com to request a special pre-registration discount.  Limited number of pre-registration coupons.

Registration Link:

TBD

 


 

Being Agile & Yet Compliant (Public or Private)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for scheduling!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

2-days onsite (4 days virtual) with group exercises, quizzes, examples, Q&A.

Instructors: Mike Russell, Ron Baerg

Next public offering: March 7 & 28, 2024

Virtual via Zoom

Registration Link:

Register Now

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.