Company: Beckman Coulter Inc.
Date of Enforcement Report: 4/29/2020
Class II
PRODUCT
Beckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Power Express Dynamic Inlet, catalog B36352.
Recall Number: Z-1741-2020
REASON
Potential exposure to biohazard. Software design problem causes excess speed and vibrations in the unload arm movement which causes caused sample splashing.
RECALLING FIRM/MANUFACTURER
Beckman Coulter Inc. on 5/17/2019. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
291 units
DISTRIBUTION
US Nationwide and International