Recall – Software issue can lead to incorrect patient positioning and dose

Company: Siemens Medical Solutions USA, Inc
Date of Enforcement Report: 3/3/2020
Class II

PRODUCT

COHERENCE Oncologist, Model Nos. 07333680 & 07351898

  • Recall Number: Z-1346-2020

REASON

A potential safety issue arises when an offset calculation is followed by a filter operation; correcting the image alignment after this sequence will result in incorrect offset values which could lead to incorrect repositioning of the patient and dose to wrong location.

RECALLING FIRM/MANUFACTURER

Siemens Medical Solutions USA, Inc on 1/16/2020. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

64 Worldwide/26 Domestic

DISTRIBUTION

US Nationwide and International

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.