Recall – Software fails to recognize new batteries

Company: Braemar Manufacturing, LLC
Date of Enforcement Report: 3/26/2020
Class II

PRODUCT

Braemar Manufacturing LLC DL900 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only.

  • Recall Number: Z-1459-2020

REASON

An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.

RECALLING FIRM/MANUFACTURER

Braemar Manufacturing, LLC on 1/22/2020. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

10582

DISTRIBUTION

US Nationwide and International

62304 Software Training Course – February 23-25, 2021

IEC 62304 and Emerging Standards and FDA Expectations for Medical Device and Health IT Software – Virtual

This very popular 3-day course provides a clear understanding of applying IEC 62304 standard for medical device software and much more. The course compares and contrasts 62304 with FDA expectations and discusses approaches for alignment. In addition, participants will learn of other relevant standards and technical reports pertinent to medical device software, HealthIT, medical mobile apps, and Software as a Medical Device (SaMD) products (e.g., 82304, 80002-1, 14971, 80001-2-x, 62366).

Participants will gain practical advice and pragmatic experience with all types of medical software. Participants will leave with a clear understanding of how to effectively and efficiently integrate 62304 compliance into their software development lifecycle (SDLC).

Register (click):  EventZilla Registration Site

Need info?  Email us at training@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.