Recall – Software fails to recognize new batteries

Company: Braemar Manufacturing, LLC
Date of Enforcement Report: 3/26/2020
Class II

PRODUCT

Braemar Manufacturing LLC DL900 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only.

  • Recall Number: Z-1459-2020

REASON

An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.

RECALLING FIRM/MANUFACTURER

Braemar Manufacturing, LLC on 1/22/2020. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

10582

DISTRIBUTION

US Nationwide and International

Remote Webmeeting Assessments

SoftwareCPR can provide remote offsite assessments to support virtual offices.  Our consultants can utilize webmeeting tools to walk your teams through assessments such as:

  • 62304 compliance
  • Regulatory submission pre-review
  • Software risk analysis
  • Cybersecurity process and validation
  • Overall ISO 14971 risk management
  • Overview of software regulation with John Murray

Email office@softwarecpr.com
for more info!

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.