Company: RAYSEARCH LABORATORIES AB Date of Enforcement Report: 1/1/2020 Class II PRODUCT RayStation stand-alone software treatment planning system, Model Numbers 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0 – Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment...
Read More
Company: GE Healthcare, LLC Date of Enforcement Report: 1/1/2020 Class II PRODUCT ApexPro Telemetry System – Product Usage: The ApexPro Telemetry System is intended for use under the direct supervision of a licensed healthcare practitioner. The system is designed to acquire and monitor physiological data for ambulating patients within a defined coverage area. The system...
Read More
Company: LivaNova USA Inc Date of Enforcement Report: 1/1/2020 Class I PRODUCT VNS Therapy, SenTiva, Model # 1000, (01)05425025750405, Rx Only for vagal nerve stimulation (VNS) therapy Recall Number: Z-0561-2020 REASON Certain Model 1000 generators (SN = 100,000) have experienced unexpected device resets, which resulted in disablement of therapy. Fourteen (14) complaints have been reported....