Recall – Radiation treatment planning system settings unintentionally result in incorrect dose calculation

Company: Elekta Inc
Date of Enforcement Report: 10/2/2019
Class II

PRODUCT

Monaco RTP System, 5.40 Unity, radiation treatment planning system Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon, electron, and carbon treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups. The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for: -contouring -image manipulation -simulation -image fusion -plan optimization -QA and plan review

  • Recall Number: Z-2560-2019

REASON

It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.

RECALLING FIRM/MANUFACTURER

Elekta Inc on 9/4/2019. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

30 units

DISTRIBUTION

US Nationwide and International

Remote Webmeeting Assessments

SoftwareCPR can provide remote offsite assessments to support virtual offices.  Our consultants can utilize webmeeting tools to walk your teams through assessments such as:

  • 62304 compliance
  • Regulatory submission pre-review
  • Software risk analysis
  • Cybersecurity process and validation
  • Overall ISO 14971 risk management
  • Overview of software regulation with John Murray

Email office@softwarecpr.com
for more info!

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.