Recall – Insertable Cardiac Monitor unable to pair with mobile app

Company: St Jude Medical Inc.
Date of Enforcement Report: 5/1/2019
Class II:

PRODUCT

St. Jude Medical Confirm Rx Insertable Cardiac Monitor, REF DM3500, Sterile. Product Usage: The St. Jude Medical Confirm Rx insertable cardiac monitor (ICM) is designed to detect arrhythmias and wirelessly transmit data to the Merlin.net Patient Care Network (PCN). It is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.

Recall Number: Z-1220-2019

REASON

The device is unable to pair with the mobile app due to the device incorrectly determining the certificate has expired.

RECALLING FIRM/MANUFACTURER

St Jude Medical Inc. on 11/19/2018. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
32 devices

DISTRIBUTION

US Nationwide

Remote Webmeeting Assessments

SoftwareCPR can provide remote offsite assessments to support virtual offices.  Our consultants can utilize webmeeting tools to walk your teams through assessments such as:

  • 62304 compliance
  • Regulatory submission pre-review
  • Software risk analysis
  • Cybersecurity process and validation
  • Overall ISO 14971 risk management
  • Overview of software regulation with John Murray

Email office@softwarecpr.com
for more info!

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.