Recall – Spine & Trauma Navigation System software error

Company: Brainlab AG
Date of Enforcement Report: 4/24/2019
Class I:

PRODUCT

Spine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 ) Intended Use The Spine & Trauma Navigation System is intended as an intraoperative image-guided localization system to enable minimally invasive surgery.

Recall Number: Z-1082-2019

REASON

In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ), instead of displaying the desirable view representation Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a crash restore or after changing between different navigation workflows during the same patient treatment.

RECALLING FIRM/MANUFACTURER

Brainlab AG on 2/27/2019. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
148 units

DISTRIBUTION

US Nationwide and International