Recall – Radiation treatment planning system software error

Company: Elekta, Inc.
Date of Enforcement Report: 2/8/2019
Class II:

PRODUCT

Monaco Radiation Treatment Planning (RTP) System

Recall Number: Z-0965-2019

REASON

If Improve Target Dose was chosen as an optimization model in a previous treatment session, Monaco will automatically use this optimization model again when proceeding with the online plan adaptation of a completion plan when it should not.

RECALLING FIRM/MANUFACTURER

Elekta, Inc. on 2/8/2019. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
14 devices

DISTRIBUTION

US Nationwide and International