Recall – Universal Viewer error leads to potential for incorrect diagnosis

Company: GE Healthcare, LLC
Date of Enforcement Report: 2/1/2019
Class II

PRODUCT

GE Healthcare Centricity Universal Viewer Breast Imaging Centricity Universal Viewer is a device that displays medical images and data from various imaging sources, and from other healthcare information sources. Medical images and data can be displayed, communicated, stored, and processed. Typical users of this system are authorized healthcare professionals. Centricity Universal Viewer is intended to assist in the viewing, analysis, diagnostic interpretation, and sharing of images and other information.

Recall Number: Z-0793-2019

REASON

When switching back & forth between multiple UV instances in the Windows taskbar, the patient images displayed on the mammo high resolution monitors may not show images of the patient selected on UV, which could result in an incorrect diagnosis.

RECALLING FIRM/MANUFACTURER

GE Healthcare, LLC, Waukesha, WI on 12/20/2018. Voluntary: Firm Initiated, recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
297 (131 US; 166 OUS)

DISTRIBUTION
U.S. Nationwide and International

Upcoming Training

62304, FDA, and Emerging Standards for Medical Device and HealthIT
Instructors:  Brian Pate, John F. Murray, Jr
Location: Sunnyvale, CA, USA
Dates:  February 4-6, 2020
Registration Link

Receive $300 discount with Premium-Individual subscription purchase (or $333 per person for Premium-Company subscription)!  Email training@softwarecpr.com
to receive discount

QSS Software Validation
Planned Instructors:  Brian Pate, John F. Murray, Jr
Location: Boston, MA, USA
Dates:  June 2-4, 2020
For info on this course, email training@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.