Recall – Universal Viewer error leads to potential for incorrect diagnosis

Company: GE Healthcare, LLC
Date of Enforcement Report: 2/1/2019
Class II

PRODUCT

GE Healthcare Centricity Universal Viewer Breast Imaging Centricity Universal Viewer is a device that displays medical images and data from various imaging sources, and from other healthcare information sources. Medical images and data can be displayed, communicated, stored, and processed. Typical users of this system are authorized healthcare professionals. Centricity Universal Viewer is intended to assist in the viewing, analysis, diagnostic interpretation, and sharing of images and other information.

Recall Number: Z-0793-2019

REASON

When switching back & forth between multiple UV instances in the Windows taskbar, the patient images displayed on the mammo high resolution monitors may not show images of the patient selected on UV, which could result in an incorrect diagnosis.

RECALLING FIRM/MANUFACTURER

GE Healthcare, LLC, Waukesha, WI on 12/20/2018. Voluntary: Firm Initiated, recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
297 (131 US; 166 OUS)

DISTRIBUTION
U.S. Nationwide and International

Software Verification

Do you have gaps with IEC 62304 compliance or with FDA expectations for software verification?  We can help you quickly remediate gaps with unit verification including unit testing, code review, and static analysis, as well as integration and system verification testing.  Call us at 781-721-2921 or email office@softwarecpr.com to start a conversation.  We can provide estimates of cost and support your development tools.

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.