Recall – UI copy error

Company: Elekta, Inc.
Date of Enforcement Report: 12/27/2018
Class II

PRODUCT

MOSAIQ Oncology Information System Version 2.64. MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery

Recall Number: Z-0686-2019

REASON

If a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup field, then the copied field will have the same isocenter as the parent and that isocenter may not be the correct isocenter for imaging when it is sent to the machine.

RECALLING FIRM/MANUFACTURER

Elekta, Inc., Atlanta, GA on 12/4/2018. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
551

DISTRIBUTION
Nationwide and Internationally

CSV Training Course

Learn FDA expectations for software validation for computer systems, quality system software, manufacturing and production process software, and engineering tools. Email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.