Recall – Software upgrade causes unexpected movement

Company: Agfa-Gevaert, N.V.
Date of Enforcement Report: 1/7/2019
Class II

PRODUCT

DX-D 600 DIRECT RADIOGRAPHY SYSTEM Product Usage: The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in hospitals, clinics and medical practices by physicists, radiographers and radiologists to make, process and view static X-Ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients.

Recall Number: Z-0692-2019

REASON

After an upgrade of the software of the Overhead Tube Crane, there were isolated cases in which the Overhead Tube Crane movement does not stop when the movement button is released. Instead of stopping while the movement button is released, the Tube Head Crane moves to the intended position.

RECALLING FIRM/MANUFACTURER

Agfa-Gevaert, N.V., Kontich, Belgium on 12/4/2018. Voluntary: Firm Initiated, recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
10

DISTRIBUTION
U.S. Nationwide