Recall – Physiomonitoring System IBP values freeze on display

Company: Invivo Corporation
Date of Enforcement Report: 11/14/2018
Class II:

PRODUCT
Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388

Recall Number: Z-0462-2019

REASON
The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard

RECALLING FIRM/MANUFACTURER
Invivo Corporation, Orlando, FL on 9/24/2018. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
76

DISTRIBUTION
Distributed to the Netherlands

Need Your Input!

We have been researching where and when to hold our next 3-day 62304/FDA training course and we want your feedback!
1) Does the location in the US for a public course matter to you … for example, west coast (e.g., San Francisco area) vs. east coast (e.g., Boston)?
2) What locations do you prefer?
3) Would you prefer just the 3-day FDA/62034 course or to have the option of a full week 3-day course followed by cybersecurity 1-day, and a usability 1-day?
4) What other countries would you want to see our course offered publicly?
5) What are best two months of the year for the course?  What are the least favored (not workable) months?
Please respond by email to

office@softwarecpr.com by Sep 25, 2019 if possible!

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.