Recall – Cl II ABL90 FLEX Analyzer

Company: Radiometer America Inc.
Date of Enforcement Report 10/3/2018
Class lI:

PRODUCT

ABL90 FLEX analyzers with software versions 3.3MR1 or below. The ABL90 FLEX Analyzer is a portable, automated Analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood. It is intended for use in a laboratory environment, near patient or point-of-care setting.
Product #: ABL90 FLEX analyzers with software versions 3.3MR1 or below.

Recall Number: Z-0010-2019

REASON
The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place: The analyzer is in ready mode Operator lifts the inlet and then closes it again, thereby initiating a rinse Immediately after lifts the inlet and quickly closes it again

RECALLING FIRM/MANUFACTURER
Radiometer America Inc. Brea, CA on 5/25/2018 Voluntary: Firm Initiated

VOLUME OF PRODUCT IN COMMERCE
12,102

DISTRIBUTION
US Distribution and Internationally to Canada

___________________________________

Upcoming Training

62304, FDA, and Emerging Standards for Medical Device and HealthIT
Instructors:  Brian Pate, John F. Murray, Jr
Location: Sunnyvale, CA, USA
Dates:  February 4-6, 2020
Registration Link

Receive $300 discount with Premium-Individual subscription purchase (or $333 per person for Premium-Company subscription)!  Email training@softwarecpr.com
to receive discount

QSS Software Validation
Planned Instructors:  Brian Pate, John F. Murray, Jr
Location: Boston, MA, USA
Dates:  June 2-4, 2020
For info on this course, email training@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.