Recall – Direct Digital Imaging System

Company: LIEBEL-FLARSHEIM COMPANY LLC
Date of Enforcement Report 8/29/2018
Class II:

PRODUCT
Liebel-Flarsheim Direct Digital Imaging System (DDIS) Liebel-Flarsheim urology systems facilitate radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation, primarily for urological applications such as functional x-ray diagnostics, endourology and minimal invasive urology/ surgery.uch as functional x-ray diagnostics, endourology and minimal invasive urology/ surgery. Recall Number: Z-2834-2018

REASON
Software issue. The difference between the display and dosimeter readings in the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the dosimeter readings.

RECALLING FIRM/MANUFACTURER
LIEBEL-FLARSHEIM COMPANY LLC, St Louis, MO on 2/27/2017. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
28

DISTRIBUTION
Nationwide

What’s New

FDA:  Exemptions, Variances, and Alternate Forms of Adverse Event Reporting for Medical Devices

Manufacturers, importers, and user facilities can request an exemption or variance from any or all of the medical device reporting requirements in 21 CFR Part 803.  Read more:  https://www.fda.gov/medical-devices/medical-device-reporting-mdr-how-report-medical-device-problems/exemptions-variances-and-alternate-forms-adverse-event-reporting-medical-devices