Company: NordicNeuroLab AS
 Date of Enforcement Report 6/13/2018
 Class II:
PRODUCT
 nordicBrainEx 2.0 provides analysis and visualization capabilities of dynamic MRI data of the brain, presenting the derived properties and parameters in a clinically useful context.. Recall Number: Z-2050-2018
REASON
 An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
RECALLING FIRM/MANUFACTURER
 NordicNeuroLab AS, Bergen Norway on 9/17/2014. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
 22 licenses
DISTRIBUTION
 Nationwide and Internationally