Company: NordicNeuroLab AS
 Date of Enforcement Report 6/13/2018
 Class II:
PRODUCT
 nordicICE v 2.3.14 nordicICE is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities. nordicICE provides both viewing and analysis capabilities of functional and dynamic imaging datasets acquired with MRI or other relevant modalities, including blood oxygen level dependent (BOLD) fMRI, diffusion weighted MRI (DWI) / fiber tracking and dynamic analysis. Recall Number: Z-2061-2018
REASON
 The bug causes BOLD activation maps to be visualized as overlays without taking the coregistration into account. Any output created from these activation maps as overlays will not be adjusted according to the co-registration.
RECALLING FIRM/MANUFACTURER
 NordicNeuroLab AS, Bergen Norway on 7/3/2015. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
 121 licenses
DISTRIBUTION
 Nationwide and Internationally