Recall – Apollo Anesthesia Machine

Company:  Draeger Medical, Inc.
Date of Enforcement Report:  6/27/2018

PRODUCT
Apollo Anesthesia Machine; Cat. no. 8605310 Product Usage:  Indicated as a continuous flow anesthesia system. The Apollo may be used for manually assisted or automatic ventilation, delivery of gases and anesthetic vapor, and monitoring of oxygen and CO2 concentration, breathing pressure, respiratory volume, and anesthetic agent concentration and identification. Recall Number: Z-2253-2018

REASON
The Draeger anesthesia device may be able to dose 100% N2O.  In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.

RECALLING FIRM/MANUFACTURER
Draeger Medical, Inc., Telford PA. on 5/24/2018. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
72

DISTRIBUTION
Nationwide

About the author

Brian is a biomedical software engineer - whatever that is! Started writing machine code for the Intel 8080 in 1983. Still enjoys designing and developing code. But probably enjoys his garden more now and watching plants grow ... and grandkids grow!

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.