Cl II THERMOCOOL SMARTTOUCH SF

Company:Biosense Webster, Inc.
Date of Enforcement Report 11/8/2017
Class lI:

PRODUCT

THERMOCOOL SMARTTOUCH SF Uni-Directional Product Codes D-1347-01-S, D-1347-02-S, D-1347-03-S
Recall Number Z-0054-2018

REASON
Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a “Map: magnetic distortion” when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy..

RECALLING FIRM/MANUFACTURER
Biosense Webster, Inc., Irwindale CA. on 9/16/2017. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
357 units

DISTRIBUTION
Nationwide and Internationally

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.