Cl II VITROS Immunodiagnostic Products

Company:ORTHO-CLINICAL DIAGNOSTICS
Date of Enforcement Report 4/192017
Class lI:

PRODUCT

VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF/Catalog 680 2156, IVD — Ortho-Clinical Diagnostics Pencoed, Bridgend CF35 5PZ, UK
Recall Number Z-1765-2017

REASON
Increased frequency of calibration failures for VITROS Immunodiagnostic Products NT-proBNP Reagent Lots 1568, 1570, 1580 and 1590, due to background signals for these affected NT-proBNP lots that have been increasing with time since release testing, resulting in higher than expected VITROS Immunodiagnostic Products NT-proBNP level 1 calibrator signals produced, leading to calibration failures.

RECALLING FIRM/MANUFACTURER
ORTHO-CLINICAL DIAGNOSTICS, FELINDRE MEADOWS ,Bridgend, United Kingdom on 3/7/2017. Voluntary: Firm Initiated recall is ongling.

VOLUME OF PRODUCT IN COMMERCE
US: 6324; Foreign: 1868

DISTRIBUTION
Nationwide and Internationally

___________________________________

Software Verification

Do you have gaps with IEC 62304 compliance or with FDA expectations for software verification?  We can help you quickly remediate gaps with unit verification including unit testing, code review, and static analysis, as well as integration and system verification testing.  Call us at 781-721-2921 or email office@softwarecpr.com to start a conversation.  We can provide estimates of cost and support your development tools.