Cl II VITROS Immunodiagnostic Products

Company:ORTHO-CLINICAL DIAGNOSTICS
Date of Enforcement Report 4/192017
Class lI:

PRODUCT

VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF/Catalog 680 2156, IVD — Ortho-Clinical Diagnostics Pencoed, Bridgend CF35 5PZ, UK
Recall Number Z-1765-2017

REASON
Increased frequency of calibration failures for VITROS Immunodiagnostic Products NT-proBNP Reagent Lots 1568, 1570, 1580 and 1590, due to background signals for these affected NT-proBNP lots that have been increasing with time since release testing, resulting in higher than expected VITROS Immunodiagnostic Products NT-proBNP level 1 calibrator signals produced, leading to calibration failures.

RECALLING FIRM/MANUFACTURER
ORTHO-CLINICAL DIAGNOSTICS, FELINDRE MEADOWS ,Bridgend, United Kingdom on 3/7/2017. Voluntary: Firm Initiated recall is ongling.

VOLUME OF PRODUCT IN COMMERCE
US: 6324; Foreign: 1868

DISTRIBUTION
Nationwide and Internationally

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.