Cl II Siemens Dimension Vista 500

Company: Siemens Healthcare Diagnostics Inc.
Date of Enforcement Report 6/1/2016
Class lI:

PRODUCT

Dimension Vista 500 Intelligent Lab System running on software versions V.3.6.1 SP1 or V.3.6.2, Device listing # D011374, All serial numbers and lots are affected. The Dimension Vista System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. Dimension Vista chemical and immunochemical applications use photometric, turbidimetric, chemiluminescence, nephelometric and integrated ion-selective multisensor technology for clinical use.
Recall Number Z-1885-2016

REASON
Discrepant patient results on Dimension Vista Intelligent Lab Systems. Siemens Healthcare confirmed a software defect which, in a very specific set of circumstances, results in the Dimension Vista System omitting an aliquot probe rinse between sample aspirations when processing tubes in Sample Racks that are front loaded on the Dimension Vista System.

RECALLING FIRM/MANUFACTURER
Siemens Healthcare Diagnostics Inc., Brookfield, CT on 3/24/2016. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
1301 units

DISTRIBUTION
Nationwide and Puerto Rico

___________________________________

CSV Training Course

Learn FDA expectations for software validation for computer systems, quality system software, manufacturing and production process software, and engineering tools. Email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.