Cl II Philips Healthcare Brilliance 64

Company: Philips Healthcare
Date of Enforcement Report 6/15/2016
Class lI:

PRODUCT

Philips Healthcare Brilliance 64 Computed Tomography X-Ray System
Recall Number Z-1713-2016

REASON
Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

RECALLING FIRM/MANUFACTURER
Philips Healthcare, Andover, MA on 9/6/2015. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
423

DISTRIBUTION
Nationwidewide

___________________________________

Upcoming Training

QSS Software Validation
Planned Instructors:  Brian Pate, John Murray
Location: Boston, MA, USA
Dates:  June 2-4, 2020
Registration Link

Receive $300 discount with Premium-Individual subscription purchase (or $333 per person for Premium-Company subscription)! 

Multi-Student Discounts as well!

Email training@softwarecpr.com
to receive discount

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.