OTS/SOUP problem, Cl II Jadak Barcode Scanner

Company: CareFusion 303, Inc.
Date of Enforcement Report 2/24/2016
Class lI:

PRODUCT

An accessory for the Pyxis Anesthesia ES system and Pyxis Anesthesia system 4000 stations. The Jadak Barcode Scanner is used to scan medication labels during refill of the Pyxis Anesthesia ES system and Pyxis Anesthesia system 4000 stations and dispensing medication for patients. Recall Number Z-0814-2016

REASON
Customers reported issues when scanning medications with the scanners. When a user scans a medication using the affected barcode scanner, information on a different medication could be communicated to the Pyxis product.

RECALLING FIRM/MANUFACTURER
CareFusion 303, Inc., San Diego CA on 1/15/2016. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
9,345 units

DISTRIBUTION
US Nationwide and the countries of Saudi Arabia, Bahrain, Australia, United Arab Emirates, Qatar, Mexico, Guam and the Bahamas.

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.