Cl II Mindray Panorama Patient Monitoring Network

Company: Mindray DS USA, Inc. dba Mindray North America
Date of Enforcement Report 2/10/2016
Class lI:

PRODUCT

Panorama Patient Monitoring Network. Software Kit 0020-00-0205-19A, View Station Software Kit 0020-00-0206-16A, Work Station Software Kit 0020-00-0207-16A, Software Kit Upgrade TECH-00-910. The Panaroma Patient Monitoring Network is intended for use in a fixed location in the healthcare facility setting as a central viewing station. It is not intended to be directly connected to patient at any time or installed in a patient’s vicinity.
Recall Number Z-0701-2016

REASON
An issue with the Panorama Central Station may cause the system to spontaneously restart. This may occur after about 49 days of continuous operation. On restart, monitoring will be lost for about 2.5 minutes. Settings and configurations are maintained..

RECALLING FIRM/MANUFACTURER
Mindray DS USA, Inc. dba Mindray North America, Mahwah, NJ on 12/15/2015. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
63 products

DISTRIBUTION
Nationwide

___________________________________

Upcoming Training

62304, FDA, and Emerging Standards for Medical Device and HealthIT
Instructors:  Brian Pate, John F. Murray, Jr
Location: Sunnyvale, CA, USA
Dates:  February 4-6, 2020
Registration Link

Receive $300 discount with Premium-Individual subscription purchase (or $333 per person for Premium-Company subscription)!  Email training@softwarecpr.com
to receive discount

QSS Software Validation
Planned Instructors:  Brian Pate, John F. Murray, Jr
Location: Boston, MA, USA
Dates:  June 2-4, 2020
For info on this course, email training@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.